Key result
New Doppler ultrasound mean frequency estimator achieves linearity within ~2% from 200 Hz to 10 kHz.
Why the study?
A practical and inexpensive analogue method for instantaneous mean frequency estimation from Doppler signals is needed for quantitative blood flow measurement.
Answers are grounded in this paper and cite their sources.
May support real-time Doppler processing in practice; leaves open in vivo validation before clinical adoption.
Ultrasonic Doppler-shifted signals from blood can be displayed in various forms using either analogue or digital processing techniques. One of the most commonly extracted parameters from the Doppler spectrum is the mean frequency envelope. This waveform can be used for the interpretation of physiological changes within man and can also be used for the estimation of volume flow rate within a vessel. This paper describes a practical implementation of an analogue, instantaneous mean frequency estimator and its use in the measurement of blood flow volume. The circuit is linear to within two per cent and operates over a frequency range of 200 Hz to 10 kHz. It can extract a velocity-time envelope from a Doppler signal within noise and performs well in a clinical environment. It is relatively inexpensive compared with digital signal processing techniques and requires minimal setting up. Simple connection to a Doppler velocimeter using phase quadrature demodulation will produce a directional output.
No takes yet. Share an insight, caveat, or question.
Beard et al. (1987) studied this question. Analogue instantaneous mean frequency estimator vs. Digital signal processing techniques was evaluated on Linearity and frequency range. An analogue instantaneous mean frequency estimator for Doppler ultrasound was developed, demonstrating linearity to within 2% over a frequency range of 200 Hz to 10 kHz.
Synapse has enriched 10 related papers. Consider them for comparative context: