Key result
Monorail balloon catheters can mimic severe coronary dissection that resolves immediately upon deflation.
Why the study?
The new Monorail balloon catheter may produce a false image of severe coronary dissection during angioplasty, which requires characterization.
Case Report (n=7)
Answers are grounded in this paper and cite their sources.
Recognition of Monorail balloon artifact avoids misdiagnosis of dissection; leaves open its prevalence and impact in larger cohorts.
During percutaneous transluminal coronary angioplasty, the appearance of persistent staining in the vessel by contrast media suggests coronary dissection. We report seven patients in whom a false image of severe coronary dissection was observed during angioplasty performed with the new Monorail balloon catheter. This image emerges at the moment of balloon inflation, is distally located to the balloon, and disappears with balloon catheter deflation. No complications were associated with the appearance of this image.
No takes yet. Share an insight, caveat, or question.
Esplugas et al. (1990) conducted a case report in Patients undergoing percutaneous transluminal coronary angioplasty (n=7). Monorail angioplasty balloon catheter was evaluated on Appearance of false image of severe coronary dissection. Use of the new Monorail balloon catheter during angioplasty produced a false image of severe coronary dissection in 7 patients, which disappeared upon deflation without complications.