Key result
Device-measured brachial systolic time intervals predict LVEF under 50% with an AUC of ~0.77.
Why the study?
It was uncertain whether systolic time intervals measured from an ankle-brachial index-form device can represent those measured from echocardiography and their diagnostic value for predicting LVEF <50%.
Cross-Sectional (n=849)
Single-blind
No
Answers are grounded in this paper and cite their sources.
Systolic time interval (STI) is an established noninvasive technique for the assessment of cardiac function. Brachial STIs can be automatically determined by an ankle-brachial index (ABI)-form device. The aims of this study are to evaluate whether the STIs measured from ABI-form device can represent those measured from echocardiography and to compare the diagnostic values of brachial and echocardiographic STIs in the prediction of left ventricular ejection fraction (LVEF) <50%. A total of 849 patients were included in the study. Brachial pre-ejection period (bPEP) and brachial ejection time (bET) were measured using an ABI-form device and pre-ejection period (PEP) and ejection time (ET) were measured from echocardiography. Agreement was assessed by correlation coefficient and Bland-Altman plot. Brachial STIs had a significant correlation with echocardiographic STIs (r = 0.644, P<0.001 for bPEP and PEP; r = 0.850, P<0.001 for bET and ET; r = 0.708, P<0.001 for bPEP/bET and PEP/ET). The disagreement between brachial and echocardiographic STIs (brachial STIs minus echocardiographic STIs) was 28.55 ms for bPEP and PEP, -4.15 ms for bET and ET and -0.11 for bPEP/bET and PEP/ET. The areas under the curve for bPEP/bET and PEP/ET in the prediction of LVEF <50% were 0.771 and 0.765, respectively. Brachial STIs were good alternatives to STIs obtained from echocardiography and also helpful in prediction of LVEF <50%. Brachial STIs automatically obtained from an ABI-form device may be helpful for evaluation of left ventricular systolic dysfunction.
No takes yet. Share an insight, caveat, or question.
Su et al. (2013) conducted a cross-sectional in Suspected cardiovascular disease (coronary artery disease, heart failure, hypertension) (n=849). Brachial systolic time intervals (ABI-form device) vs. Echocardiographic systolic time intervals was evaluated on Correlation between brachial and echocardiographic ratio of pre-ejection period to ejection time (bPEP/bET vs PEP/ET) (r = 0.708, p=<0.001). Brachial systolic time intervals measured by an ABI-form device correlated with echocardiographic measurements and effectively predicted LVEF <50% with an AUC of 0.771.
Synapse has enriched 10 related papers. Consider them for comparative context: