Key result
Micra AV requires tailored mechanical atrial sensing programming to optimize AV synchrony.
Why the study?
The Micra AV leadless pacemaker provides AV synchrony via a novel mechanical atrial contraction algorithm that has limitations and requires specific programming and troubleshooting guidance.
Micra leadless pacemaker has progressed from a single chamber pacemaker that can deliver VVIR pacing to a pacing device that can provide atrio-ventricular (AV) synchrony via a unique pacing algorithm that relies on identifying mechanical atrial contraction. This novel algorithm has its own limitations and intricacies. In this paper, we review this algorithm, suggest steps for troubleshooting and programming these devices and provide clinical examples of Micra AV cases that required changes in programming for adequate tracking of atrial activity.
No takes yet. Share an insight, caveat, or question.
El‐Chami et al. (2020) conducted a review in Brady-arrhythmias / Complete heart block. Micra AV leadless pacemaker programming and troubleshooting was evaluated. The Micra AV leadless pacemaker relies on mechanical atrial contraction detection to provide AV synchrony, often requiring manual programming of A3 and A4 thresholds for optimal function.