Background & Aims:We present an analytical formula (AF) to calculate the total scatter factor of the MLC-shaped small radiation beams. Materials & Methods:Pinpoint, Semiflex, and Farmer ion chambers were used for MLC/Jaw-shaped fields extending from (10x10 cm 2 ) down to (1x1 cm 2 ).The total scatters factor, beam profiles, and penumbra was investigated for MLC/Jaw-shaped beams.Results: We found that the total scatters factor and the penumbra was clearly higher for MLC-shaped beams due to the MLC leakage, which becomes more and more significant for the small fields.The Analytical Formula was introduced as a least-square fit equation and verified against to measure data.The maximum deviation at small field size (< 4 cm 2 ) was within 3.6% and 43.9% from both MLC and Jaw-shaped beams, respectively.Conclusion: Our analytical formula shows a very good agreement with measured data at a large field size (> 4 cm 2 ), and had small deviation with MLC-shaped small field size, emphasizing that our analytical formula is valid and reasonably good for the total scatters factor calculation of the MLC-shaped beams, especially the ultra-small field size.