Key result
Trefoil balloon valvuloplasty reduces transvalvar aortic gradient by ~30 mm Hg in congenital aortic stenosis.
Why the study?
A new balloon design for aortic valvuloplasty was developed to maintain blood flow during dilation in congenital aortic stenosis.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
A new balloon for valvuloplasty (Trefoil balloon) was successfully used in a 12 year old boy with congenital aortic stenosis. The Trefoil balloon consists of three angioplasty balloons mounted in parallel on a single shaft. When they are inflated simultaneously they produce a rosette that allows blood flow to continue through the valve. The aortic pressure only dropped to 75 mm Hg during a 15 s balloon dilatation at 4 bar (400 kPa). The procedure reduced the transvalvar gradient from 45 to 15 mm Hg.
No takes yet. Share an insight, caveat, or question.
Meier et al. (1986) conducted a case report in congenital aortic stenosis (n=1). Trefoil balloon for valvuloplasty was evaluated on transvalvar gradient. Trefoil balloon valvuloplasty successfully reduced the transvalvar gradient from 45 to 15 mm Hg in a 12-year-old boy with congenital aortic stenosis.