A non-enzymatic electrochemical glucose sensor was developed based on Ni@C/NiO–Ni(OH) 2 heterostructures integrated into a biomass-derived carbon aerogel (CA) via a simple, cost-effective approach. The three-dimensional porous CA, obtained from sugarcane bagasse, provides a conductive framework with high surface area for uniform dispersion of active species. During carbonization, Ni nanoparticles catalyze the formation of a thin carbonaceous layer containing a moderate density of defects, while partially exposed Ni is converted into NiO–Ni(OH) 2 , generating abundant Ni(II)/Ni(III) redox sites. This heterostructure enhances electron transfer and increases the accessibility of catalytic centers, resulting in improved electrocatalytic performance toward glucose oxidation in alkaline media compared with conventional Ni-based systems. The sensor exhibits a linear range of 1–13 mM, a sensitivity of 102.1 µA mM –1 cm −2 , and a low detection limit of 0.03 mM, along with good selectivity, reproducibility, and stability. Practical applicability was confirmed by accurate glucose determination in commercial beverages, yielding recoveries of 96.25–105.00% with relative standard deviations below 5%.