In light of the recent article by Spencer et al. (1) as well as the editorial by Spencer (2), we feel obliged to clarify a few points concerning the performance of our thyroglobulin (Tg) immunoassay and make some more general comments. In Fig. 1 of ref. 1, the legend reads “Serum Tg values measured by six different Tg methods.” The six assays tested are clearly indicated in the legend but only four of them appear in the Figure. The one from our group, Tg IRMA Pasteur, is missing. Surprisingly, the “relative sensitivities of the five different Tg methods as judged by the response in signal cpm,” shown in Fig. 2 of ref. 1, are expressed as absolute signals given by the bound (B) antibody tracer (background subtracted). When antibodies are labeled with radioisotopes that have a relatively short half-life, the results should preferably be expressed as the bound fraction (B/B0) to take into account radioactive decay. In our case (the Sanofi Diagnostics Pasteur test has a 125I label), the difference between the two modes of representation is particularly striking because the assay was performed approximately at the expiration date of the tracer. All evaluations of our Tg immunoassay, without fail, underlined the very high analytical sensitivity of our assay due to the use of a selected combination of antibodies.