Key result
Intentional disc separation enables successful, complication-free ASD occluder deployment across three cases.
Why the study?
Imaging assessment of petal orientation and septal apposition during GORE CARDIOFORM ASD Occluder deployment is challenging due to disc overlap obscuring visualization.
Case Report (n=3)
Answers are grounded in this paper and cite their sources.
An intentional disc separation maneuver improves diagnostic confidence during transcatheter closure of complex atrial septal defects by facilitating clear visualization of petal orientation using transesophageal echocardiography and fluoroscopy.
BACKGROUND: Transcatheter closure of atrial septal defects with the GORE CARDIOFORM atrial septal defect (ASD) Occluder relies on imaging as mechanical stability testing is not recommended. However, disc overlap often obscures petal orientation and septal apposition. CASE SUMMARY: We describe 3 patients with complex ASD morphologies, including deficient rims and a floppy septum, successfully treated using an "intentional disc separation maneuver" before locking. By gently retracting the delivery slider while maintaining position, a temporary gap was created between the discs, facilitating complementary assessment of petal orientation by transesophageal echocardiography and fluoroscopy. Strictly an imaging aid rather than a mechanical test, this technique enabled accurate assessment in challenging anatomies, such as multiple rim deficiencies, absent aortic rim, and atrial septal aneurysm. Confident deployment was achieved in all cases without complications or residual shunting. CONCLUSIONS: The "intentional disc separation maneuver" improves diagnostic confidence during conformable ASD occluder deployment in complex cases.
No takes yet. Share an insight, caveat, or question.
Ohashi et al. (2026) conducted a case report in Atrial septal defects with complex morphologies (n=3). Intentional disc separation maneuver was evaluated on Confident deployment without complications or residual shunting. An intentional disc separation maneuver enabled accurate imaging assessment and confident ASD occluder deployment without complications or residual shunting in all 3 patients.
Synapse has enriched 10 related papers. Consider them for comparative context: