Key result
Temperature-based rate-adaptive leadless pacemakers can cause wide-complex tachycardia (~130 beats/min) during cardiopulmonary bypass rewarming, resolving when rate response is disabled.
Answers are grounded in this paper and cite their sources.
For patients with temperature-responsive leadless pacemakers undergoing cardiac surgery with cardiopulmonary bypass, preoperative device interrogation and temporary deactivation of rate response should be strongly considered to prevent rewarming-induced intraoperative tachycardia.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Binaya Basyal (2026) conducted an editorial in Intraoperative tachycardia during cardiopulmonary bypass rewarming in a patient with a temperature-responsive leadless pacemaker (n=1). Central venous temperature-based rate-adaptive pacing (Abbott AVEIR VR leadless pacemaker) during cardiopulmonary bypass rewarming was evaluated on Heart rate during cardiopulmonary bypass rewarming. Temperature-based rate-adaptive leadless pacemakers can cause wide-complex tachycardia (~130 beats/min) during cardiopulmonary bypass rewarming, resolving when rate response is disabled.