INTRODUCTION: High-performance whole-body gradient systems show promise for breast diffusion-weighted imaging, but their performance is constrained by physiological limits. Although local nonlinear gradient coils can provide stronger gradients while remaining within these limits, they often exhibit a substantial reduction in the gradient amplitude of the z-component of the magnetic field near the chest wall. METHODS: We introduced an optimization framework incorporating a constraint on the anterior-posterior variation in the encoding gradient amplitude. A figure of merit combining coil efficiency and minimum wire width was defined to assess performance. A prototype was constructed to validate the design methodology. RESULTS: The optimized coil achieved a 2.35-fold efficiency increase over standard linear coils. Compared to previous nonlinear designs, the new constraint reduced spatial variation by 35.7% and improved minimum efficiency near the chest wall by 2.6-fold. Experimental field maps matched simulations with errors under 8%. DISCUSSION: The proposed method effectively mitigates the trade-off between gradient strength and spatial uniformity along anterior-posterior direction. By enhancing performance in the posterior breast region, the design addresses a critical limitation of previous local coils. CONCLUSION: This framework enables the development of high-performance, robust local gradient coils, facilitating the clinical implementation of advanced DWI protocols for breast cancer screening.
No takes yet. Share an insight, caveat, or question.
Jia et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: