Answers are grounded in this paper and cite their sources.
BACKGROUND: A prerequisite for successful robot-assisted neurosurgery is to use a hand-controller matched with characteristics of real robotic microsurgery. This study reports quantified data pertaining to the required workspace and exerted forces of surgical tools during robot-assisted microsurgery. METHODS: A surgeon conducted four operations in which the neuroArm surgical system, an image-guided computer-assisted manipulator specifically designed to perform robot-assisted neurosurgery, was employed to surgically remove brain tumors. The position, orientation, and exerted force of surgical tools were measured during operations. RESULTS: Workspace of the neuroArm manipulators, for the cases studied, was 60×60×60 mm(3) while it offered orientation ranges of 103°, 62° and 112°. The surgical tools exerted a maximum force of 1.86 N with frequency band of less than 20 Hz. CONCLUSIONS: This data provides important information specific to neurosurgery that can be used to select among commercially available, or further design a customized, haptic hand-controller for robot-assisted neurosurgical systems.
No takes yet. Share an insight, caveat, or question.
Maddahi et al. (2015) studied this question.