Key result
Laser-etched parallel groove patterning on mechanical valve leaflets increases mean transvalvular pressure across flow rates.
Why the study?
The effects of the micro-structure of pyrolytic carbon on the hydrodynamic performance of artificial heart valves were not well understood.
Answers are grounded in this paper and cite their sources.
OBJECTIVE: We aimed to elucidate the effects of the micro-structure of the pyrolytic carbon for artificial heart valves on its hydrodynamic performance. METHODS: Bileaflet mechanical valves of GKS 23 and 29 A were randomly selected. According to ISO5840, mean transvalvular pressure (MPG), regurgitation fraction (RF), and effective orifice area (EOA) of valve were assessed. Then, parallel-groove pattern was constructed by laser etching on leaflet surface, and the valves were subjected again to the same test. RESULTS: Compared with before patterning at 2, 3.5, 5, and 7 L/min, the MPG of the valves in two specifications were higher, the EOA was larger in 23 A, but smaller in 29 A, and the RF was contrary to EOA. At 5 L/min, the RF in both specifications was lower after etching at 45 bpm. At 70 bpm however, the RF in 23 A decreased, in 29 A increased. CONCLUSION: The parallel-groove pattern on leaflet surface affected the hemodynamic performance of the valve prostheses.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2023) studied Artificial heart valves. Parallel-groove pattern by laser etching on leaflet surface vs. Before patterning was evaluated on Hydrodynamic performance (mean transvalvular pressure, regurgitation fraction, and effective orifice area). Parallel-groove patterning by laser etching on mechanical heart valve leaflets altered hemodynamic performance, including higher mean transvalvular pressure across multiple flow rates.
Synapse has enriched 10 related papers. Consider them for comparative context: