Key result
Wrist cuff occlusion yields similar mean VRI to upper-arm occlusion but shows only moderate individual agreement.
Why the study?
Whether wrist cuff occlusion provides comparable vascular reactivity index measurements to standard upper-arm cuff occlusion in digital thermal monitoring was unknown.
Does wrist cuff occlusion provide comparable vascular reactivity index (VRI) measurements to standard upper-arm cuff occlusion in healthy adults?
Cross-Sectional (n=48)
Randomized cross-over sequence
No
Does wrist cuff occlusion provide comparable vascular reactivity index (VRI) measurements to standard upper-arm cuff occlusion in healthy adults?
Mean Difference: 0.06 (95% CI -0.04–0.15)
Absolute Event Rate: 1.76% vs 1.7%
p-value: p=0.25
Wrist cuff occlusion provides similar mean Vascular Reactivity Index values to upper-arm occlusion but has only moderate individual agreement, indicating the sites are not interchangeable for individual tracking.
Wrist occlusion shows insufficient individual agreement for clinical VRI interchangeability; leaves open its utility pending larger validation studies.
Background Digital Thermal Monitoring (DTM) is an automated method for assessing vascular reactivity using the vascular reactivity index (VRI), a temperature-derived surrogate marker associated with endothelial function. Standard DTM requires five minutes of supra-systolic upper-arm cuff occlusion, which may cause discomfort and limit feasibility in some individuals. Whether wrist cuff occlusion provides comparable VRI measurements remains unknown. Methods In this method-comparison study, 48 healthy adults underwent DTM assessment using both upper-arm and wrist cuff occlusion in randomized order during the same visit. Agreement between the two occlusion sites for VRI, pulse rate (PR), and blood pressure (BP) was evaluated using paired comparisons and agreement analyses. Results Mean VRI was 1.70 ± 0.33 with upper-arm cuff occlusion and 1.76 ± 0.34 with wrist cuff occlusion, with no significant difference between sites (mean difference 0.06, 95% CI −0.04 to 0.15; p=0.25) and no significant proportional bias. Agreement for VRI was moderate (ICC = 0.53, 95% CI 0.29–0.70). In contrast, wrist measurements overestimated systolic BP by 10.5 mmHg (95% CI 7.3–13.7; p<0.001) and diastolic BP by 10 mmHg (95% CI 7.1–12.9; p<0.001), with poor agreement (ICC = 0.29, 95% CI −0.05 to 0.56 and ICC = 0.18, 95% CI −0.07 to 0.43, respectively). Conclusions Wrist cuff occlusion provides similar mean VRI values to upper-arm occlusion but demonstrates only moderate agreement at the individual level. Further studies with larger and more diverse populations, including repeated measurements, are required assessment before wrist and upper-arm cuff occlusion can be considered interchangeable for individual assessment.
No takes yet. Share an insight, caveat, or question.
Mohammadi et al. (2026) conducted a cross-sectional in Healthy adults (n=48). Wrist cuff occlusion vs. Upper-arm cuff occlusion was evaluated on Vascular Reactivity Index (VRI) (mean difference 0.06, 95% CI -0.04 to 0.15, p=0.25). Wrist cuff occlusion provided similar mean VRI values to upper-arm occlusion (mean difference 0.06, p=0.25) but demonstrated only moderate absolute agreement at the individual level (ICC 0.53).