Key result
rtPA lysis for HVAD thrombosis is followed by acute outflow graft occlusion in ~25% of patients.
Why the study?
Acute outflow graft occlusions (OGO) associated with lysis therapy for LVAD intra-pump thrombosis were repeatedly observed in clinical routine but not well understood.
Observational (n=962)
Substantial outflow graft occlusion risk after rtPA for HVAD thrombosis warrants caution; leaves open need for prospective validation of flow-based selection.
Lysis therapy is an established treatment option for intra-pump thrombosis of left ventricular assist devices (LVADs). In clinical routine, we observed repeated cases of acute outflow graft occlusions (OGO) associated with lysis therapy with need for urgent intervention. The aim of this investigation was to gain understanding of this observation. We screened data of 962 HeartWare ventricular assist device (HVAD) patients. One hundred twenty (13.8%) had intra-pump thromboses; 58 were treated with recombinant tissue-type plasminogen activator (rtPA). Mean age was 53.0 ± 11.1 years; 84.9% were male. In 13 (24.5%) patients, OGO occurred following rtPA-lysis. These patients showed an increase in left ventricular function (18.45% ± 12.62% to 27.73% ± 10.57%; p = 0.056), more frequent 1:1 aortic valve opening (OGO+: +36.4%; OGO-: +7.4%; p = 0.026), a decrease in LVAD pulsatility within 12 months prior intra-pump thrombosis (OGO+: -0.8 L/min [interquartile range {IQR}, -1.4 to -0.4 L/min]; OGO-: -0.3 L/min [IQR, -0.9 to 0.1 L/min]; p = 0.038) and lower HVAD flows at admission (OGO+: 6.7 L/min [IQR, 6.1-7.4 L/min]; OGO-: 8.3 L/min [IQR, 6.9-9.3 L/min]; p = 0.013), indicating a subclinical OGO prior intra-pump thrombosis. There were no differences in implantation techniques, blood parameters, and lysis strategy. Subclinical OGO represented a major risk factor for acute OGO following rtPA lysis therapy. We here propose an algorithm for risk stratification and dealing with patients presenting this first-described complication. Further research is required to confirm our results and decipher the underlying pathomechanism. http://links.lww.com/ASAIO/B97.
No takes yet. Share an insight, caveat, or question.
Hoermandinger et al. (2023) conducted an observational in Left ventricular assist device (LVAD) intra-pump thrombosis (n=962). rtPA lysis therapy was evaluated on Acute outflow graft occlusion (OGO) following rtPA-lysis. Acute outflow graft occlusion occurred in 24.5% of patients following rtPA lysis for intra-pump thrombosis, with lower admission HVAD flows (6.7 vs 8.3 L/min; p=0.013) being a major risk factor.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: