Key result
Micra leadless pacemakers maintain full functionality without malfunction under high-dose photon radiotherapy up to 200 Gy.
Why the study?
Data on the safety and performance of Micra leadless cardiac pacemakers during clinically relevant radiotherapy exposures are limited.
Does photon radiotherapy exposure induce device malfunction or battery depletion in leadless cardiac pacemakers?
Does photon radiotherapy exposure induce device malfunction or battery depletion in leadless cardiac pacemakers?
Leadless cardiac pacemakers maintained full stability and function during clinically relevant and extreme photon radiotherapy regimens up to 200 Gy in a phantom model.
Answers are grounded in this paper and cite their sources.
Supports Micra use in radiotherapy; extends bench data but leaves clinical safety open.
Background: With the growing use of leadless cardiac pacemakers (LCPs), radiotherapy (RT) in these patients is increasingly common. Radiation may induce LCP malfunction, yet data on Micra™ safety during RT are limited. Aims: This study aimed to assess Micra™ performance during clinically relevant photon RT exposures. Methods: Three Micra™ LCPs were placed in an anthropomorphic phantom and irradiated with 6 MV flattened (FF), 6 MV flattening-filter-free (FFF), and 10 MV FFF photon beams. Devices were positioned 0.5 cm outside the field (scatter) or at isocenter (in-field). Fractionation regimens reflected typical clinical protocols (40.05 Gy/15 fractions, 26 Gy/5 fractions, 60 Gy/8 fractions, 27 Gy/3 fractions), including cumulative doses up to 200 Gy. Delivered doses were verified with ionization chamber measurements. After each fraction (or dose step in the 200 Gy series), device function and battery voltage were checked using a Medtronic 2090 programmer for resets, errors, or mode changes. Results: Across all conditions, including single-fraction doses ≥5 Gy, dose rates up to 2400 MU/min, and cumulative in-field doses of 200 Gy, no malfunctions, resets, or error messages occurred. Battery voltage variation was ≤0.02 V (≤0.6% of baseline) without progressive decline. Devices with partially depleted batteries remained equally stable. Conclusions: Micra™ LCPs retained full functionality during RT, even under extreme high dose and high-dose-rate conditions. Although limited by the small sample size and lack of pacing parameter testing, these results indicate that Micra™ devices may safely remain in situ during a wide range of photon RT treatments with proper dosimetric planning and monitoring.
No takes yet. Share an insight, caveat, or question.
Kamińska et al. (2026) studied Radiotherapy safety of leadless cardiac pacemakers (n=3). Photon beam radiotherapy was evaluated on Observed malfunction rate (resets, error messages, or mode changes) (95% CI 0.0-5.4). Micra™ leadless pacemakers exposed to clinically relevant and extreme photon radiotherapy protocols up to 200 Gy retained full functionality without any malfunctions, resets, or mode changes.
Synapse has enriched 10 related papers. Consider them for comparative context: