Key result
Photon radiotherapy causes 0% device malfunctions or resets in leadless and transvenous pacemakers in vitro.
Why the study?
Current recommendations regarding radiation therapy effects on cardiac implantable electronic devices are largely based on legacy transvenous pacemakers, and data on leadless pacemakers remain limited.
Does photon-based stereotactic body radiation therapy and supraclinical dose escalation cause device malfunction in contemporary leadless and transvenous pacemakers tested in an in vitro phantom model?
Does photon-based stereotactic body radiation therapy and supraclinical dose escalation cause device malfunction in contemporary leadless and transvenous pacemakers tested in an in vitro phantom model?
Contemporary leadless and transvenous pacemakers demonstrate high in vitro resistance to photon-based radiotherapy, maintaining normal function even at supraclinical radiation doses far exceeding clinical exposure.
Answers are grounded in this paper and cite their sources.
Contemporary pacemakers tolerate high-dose photons in vitro; hypothesis-generating, should not yet change clinical practice.
BACKGROUND Radiation therapy (RT) is widely used contemporarily, yet concerns persist regarding its potential impact on cardiac implantable electronic devices. Current recommendations are largely based on legacy transvenous pacemakers (tvPM), and data on leadless pacemakers (LPMs) remain limited. OBJECTIVES This study aimed to evaluate the RT tolerance of LPMs and contemporary tvPM under stereotactic body radiation therapy (SBRT) and supraclinical dose escalation. METHODS Sixteen devices (11 LPMs and 5 tvPMs) were tested in vitro with a thoracic anthropomorphic phantom. Clinically simulated SBRT was delivered using 6 MV photon beams (tumor dose 54 Gy), to cumulative device doses of 1.15-1.29 Gy for LPMs and 1.81 Gy for tvPM. A subset of LPMs additionally underwent robotic stereotactic radiosurgery. Subsequently, open-field irradiation with dose escalation was performed up to 200-330 Gy, with one LPM exposed to 1,055 Gy. Devices were interrogated before and after each irradiation and at late follow-up. RESULTS No device resets, pacing mode changes, loss of programmability, or integrity alerts were observed during and after SBRT, robotic, or open-field irradiation up to 200-1,055 Gy. Battery voltage remained stable in all systems except for one LPM exposed to 1,055 Gy, in which a late reduction in estimated battery capacity occurred. All telemetry and interrogations were undisturbed. CONCLUSIONS In an anthropomorphic phantom model, modern LPMs and tvPMs demonstrated resistance to RT during clinically simulated SBRT and at doses far exceeding those clinically encountered. These findings suggest that radiation sensitivity of contemporary pacemakers may be lower than in legacy devices.
No takes yet. Share an insight, caveat, or question.
Tajstra et al. (2026) studied Radiation therapy exposure on leadless and transvenous pacemakers (n=16). Photon-based radiation therapy (SBRT, robotic stereotactic radiosurgery, and open-field irradiation) vs. Baseline pre-irradiation device function was evaluated on Device resets, pacing mode changes, loss of programmability, or integrity alerts. Photon-based radiotherapy caused 0% device resets, mode changes, or programmability loss in 11 leadless and 5 transvenous pacemakers tested in vitro up to 200-1,055 Gy.