Key result
Coagulation-mode electrosurgical energy enables safe lead tip separation without collateral injury during extraction.
Why the study?
The increasing use of cardiac implantable electronic devices has led to more transvenous lead extractions, with challenges in safe lead tip removal due to fibrotic encapsulation and thin myocardial walls.
Does electrosurgical energy in coagulation mode safely facilitate pacemaker lead tip extraction without collateral tissue damage compared to cut mode in a preclinical model?
Does electrosurgical energy in coagulation mode safely facilitate pacemaker lead tip extraction without collateral tissue damage compared to cut mode in a preclinical model?
Unipolar electrosurgical energy in coagulation mode with a fulguration setting provides a controlled and safe adjunct to manual traction for extracting chronically implanted pacemaker leads.
Hypothesis-generating for coagulation-mode electrosurgery in lead extraction; leaves open prospective human validation.
The increasing use of cardiac implantable electronic devices has led to a rise in transvenous lead extractions (TLEs). The most common indications for TLE are infections and lead malfunction. Over time, leads develop fibrotic encapsulation that makes extraction difficult and increases the risk of myocardial injury. Given the thin myocardial wall, particularly in the right atrial appendage and right ventricular apex, the use of advanced powered tools near the lead tip is contraindicated, posing additional challenges. Electrosurgical energy (EE) has been explored as an adjunct to manual extraction to facilitate safe lead tip removal. This study compares the efficacy and safety of cut and coagulation modes of EE using a porcine tissue model and reports its clinical application in a human case. In the porcine model, the cut mode achieved rapid separation of the lead tip but resulted in deeper tissue penetration. In contrast, the coagulation mode (fulguration setting) enabled controlled dissection with minimal depth of penetration without any collateral injury. The application of the coagulation mode during an actual lead extraction with a dwell time of 8 years in a human patient resulted in complete removal with no postoperative complications. These findings suggest that the coagulation mode of EE offers a safer and more controlled technique for lead tip extraction.
No takes yet. Share an insight, caveat, or question.
Reddy et al. (2026) studied Chronically implanted pacemaker/defibrillator lead requiring extraction (porcine model and clinical case) (n=1). Electrosurgical energy in coagulation mode vs. Electrosurgical energy in cut mode (preclinical) / manual traction (clinical) was evaluated on Efficacy of lead separation and safety regarding collateral tissue injury. Electrosurgical energy in coagulation mode enabled safe and controlled lead tip separation without collateral injury in a porcine model, translating to a successful and uncomplicated extraction in a human patient.