Key result
Flexible Sphere-9 catheter tip unexpectedly crosses into right atrium through small interatrial defect.
Why the study?
Unexpected right atrial crossing of the Sphere-9 catheter occurred despite stable left atrial sheath position, highlighting a procedural uncertainty.
Case Report (n=1)
Physicians using the Affera platform should be aware of potential unexpected right atrial crossing during left atrial septal mapping to avoid inadvertent ablation outside the target chamber.
Operators should remain vigilant for unintended right atrial crossing with the Sphere-9 during left atrial access; this case leaves open incidence and prevention.
The Sphere-9 catheter unexpectedly crossed into the right atrium despite stable left atrial sheath position. A perpendicular, sheath-supported approach may have facilitated passage of the flexible lattice tip through a small interatrial communication while the sheath remained within the left atrium.
No takes yet. Share an insight, caveat, or question.
Ikenaga et al. (2026) conducted a case report in Recurrent atrial fibrillation (n=1). Affera Sphere-9 catheter was evaluated. The flexible lattice tip of the Sphere-9 catheter unexpectedly crossed into the right atrium through a small interatrial communication while the sheath remained stable in the left atrium.