Key result
Totally thoracoscopic ablation maintains sinus rhythm in ~87% of patients at 1 year.
Why the study?
Outcome data for the Gemini S bipolar radiofrequency clamp in thoracoscopic AF ablation were almost absent, limited to a three-patient case series.
Does totally thoracoscopic atrial fibrillation ablation with the Gemini S bipolar clamp prevent atrial fibrillation recurrence in patients with atrial fibrillation?
Cohort (n=23)
No
Does totally thoracoscopic atrial fibrillation ablation with the Gemini S bipolar clamp prevent atrial fibrillation recurrence in patients with atrial fibrillation?
In a small single-center cohort, totally thoracoscopic AF ablation with the Gemini S bipolar clamp achieved an 87.0% freedom from recurrence at 1 year, though 13% of cases required conversion to sternotomy for bleeding.
Provides initial cohort data on Gemini S clamp outcomes; leaves open need for randomized confirmation before wider adoption.
Background Outcome data for the Gemini S bipolar radiofrequency clamp are almost absent; the only indexed report being a three‐patient case series. We describe 1‐year outcomes in a consecutive cohort. Methods We retrospectively analyzed all consecutive patients treated with the Gemini S clamp between January 2021 and August 2024. Epicardial box isolation was combined with left atrial appendage (LAA) exclusion; completeness was confirmed by entry and exit block on epicardial three‐dimensional mapping. The primary outcome was freedom from AF recurrence at 12 months, on and off antiarrhythmic drugs (AADs) after a 90‐day blanking period. Results Twenty‐three patients were included (mean age 64.4 ± 8.0 years; 82.6% male). AF was paroxysmal in 34.8%, persistent in 26.1%, and long‐standing persistent in 39.1%; 60.9% had prior catheter ablation and 69.6% had failed or were intolerant to a Class I or III agent. A complete block‐verified box lesion was achieved in 95.7% and LAA exclusion in 91.3%. Three patients (13.0%) required conversion to sternotomy for bleeding, exceeding larger series. There was no death, stroke, phrenic nerve palsy, or pacemaker implantation. At 1 year, 87.0% (95% CI 67.9–95.5) were in sinus rhythm, all off Class I/III AADs. Conclusions Sinus rhythm was achieved in 87.0% of patients at 1 year. As the largest reported experience with this device, the series provides device‐level evidence where little existed, but it is descriptive, not powered for comparison, and the conversion rate needs confirmation.
No takes yet. Share an insight, caveat, or question.
Gruszczyński et al. (2026) conducted a cohort in Atrial fibrillation (n=23). Totally thoracoscopic atrial fibrillation ablation with the Gemini S bipolar clamp was evaluated on Freedom from AF recurrence at 12 months, on and off antiarrhythmic drugs after a 90-day blanking period (95% CI 67.9-95.5). Totally thoracoscopic atrial fibrillation ablation using the Gemini S bipolar clamp resulted in 87.0% (95% CI 67.9-95.5) of patients remaining in sinus rhythm at 1 year.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: