Key result
Mason-Likar placement shifts the Wilson central terminal anteriorly by ~72 mm, distorting limb-lead directions.
Why the study?
The Mason-Likar modification is standard for exercise and ambulatory ECG but distorts recordings relative to standard distal limb electrode placement, requiring quantification of this distortion.
Does Mason-Likar electrode placement distort the ECG recording relative to standard distal placement in patient-specific models?
Does Mason-Likar electrode placement distort the ECG recording relative to standard distal placement in patient-specific models?
Mason-Likar electrode placement systematically distorts the frontal axis, reweights limb-lead amplitudes, and alters chest lead signals, suggesting that important ECG findings using this placement should be confirmed on a standard 12-lead.
Answers are grounded in this paper and cite their sources.
Mason-Likar distorts WCT position and limb axes in models; extends modeling data but leaves open effects on clinical ECG interpretation.
BACKGROUND The Mason-Likar (ML) modification is standard for exercise and ambulatory ECG but distorts the recording relative to standard distal placement. This was quantified against a proximal-limb (Lund) reference. METHODS In 12 patient-specific torso-heart models, ML and Lund limb electrodes and precordials V1-V9 were positioned. We derived the Wilson-central-terminal (WCT) displacement, each limb lead's frontal-plane direction and antero-posterior (AP) fraction, and - in the one subject with a validated boundary-element-method (BEM) forward model - the signal-level limb amplitudes and the offset added to the chest leads across 32 beats. RESULTS Across 12 models ML displaced the WCT 72.4 ± 15.9 mm anteriorly (range 51.9 to 96.3 mm). Limb‑lead directions departed from textbook (II +42°, aVF +65°), and the inferior lead axes tilted out of the frontal plane, acquiring anterior forces (AP-fraction roughly doubling). In the one subject with a forward model (32 paced beats), ML amplified II, III and aVF (×1.37, ×1.40, ×1.47) and attenuated I (×0.64) and aVR (×0.79), placing a + 45° QRS at about +67°, and added one offset to every chest lead (median 0.35 mV, range 0.10-0.59 mV), its relative weight reaching 90-140% in V7-V9. CONCLUSIONS ML modestly tilts the measured frontal axis, reweights limb‑lead amplitudes with the inferior leads acquiring anterior forces, and adds a common offset to every chest lead. The geometry rests on 12 models, the signal-level magnitudes on one subject. Confirm important inferior, lateral or precordial findings on a standard 12‑lead.
No takes yet. Share an insight, caveat, or question.
José Nunes de Alencar (2026) studied this question. Mason-Likar electrode placement vs. Proximal-limb (Lund) reference was evaluated on Wilson-central-terminal (WCT) displacement. Mason-Likar electrode placement displaced the Wilson-central-terminal 72.4 ± 15.9 mm anteriorly and distorted limb-lead directions compared to the standard proximal-limb reference.
Synapse has enriched 10 related papers. Consider them for comparative context: