Key result
Programming four ATP sequences linked to ~98% success in terminating slow VT versus default settings.
Why the study?
The efficacy of different antitachycardia pacing (ATP) programs in patients with ICDs was evaluated due to uncertainty about optimal device programming for VT treatment.
Observational (n=18,679)
S=4 ATP programming may optimize slow VT termination; hypothesis-generating and requires randomized confirmation before practice change.
BACKGROUND: Antitachycardia pacing (ATP) is an effective treatment for ventricular tachycardia (VT). We evaluated the efficacy of different ATP programs based on a large remote monitoring data set from patients with implantable cardioverter-defibrillators (ICDs). METHODS: A dataset from 18,679 ICD patients was used to evaluate the first delivered ATP treatment. We considered all device programs that were used for at least 50 patients, leaving us with 7 different programs and a total of 32,045 episodes. We used the two-proportions z-test (α = 0.01) to compare the probability of success and the probability for acceleration in each group with the corresponding values of the default setting. RESULTS: Overall, the first ATP treatment terminated in 78.4%-97.5% of episodes with slow VT and 81.5%-91.1% of episodes with fast VT. The default setting of the ATP programs with the number of sequences S = 3 was applied to treat 30.1% of the slow and 36.6% of the fast episodes. Reducing the maximum number of sequences to S = 2 decreased the success rate for slow VT (P < 0.0001, h = 0.38), while the setting S = 4 resulted in the highest success rate of 97.5% (P < 0.0001, h = 0.27). CONCLUSION: While the default programs performed well, we found that increasing the number of sequences from 3 to 4 was a promising option to improve the overall ATP performance.
No takes yet. Share an insight, caveat, or question.
Shakibfar et al. (2019) conducted an observational in Ventricular tachycardia (n=18,679). Anti-tachycardia pacing (ATP) programming with 4 sequences (S=4) vs. Default ATP programming (S=3 sequences) was evaluated on Successful termination of slow ventricular tachycardia by first ATP treatment (Cohen's h = 0.27, p=<0.0001). Programming anti-tachycardia pacing devices with a maximum of 4 sequences (S=4) achieved the highest success rate of 97.5% for terminating slow ventricular tachycardia compared to default programming.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: