Key result
Antitachycardia pacing successfully terminated 80% of spontaneous ventricular tachycardia episodes, with the first sequence being the most effective (59% success).
Why the study?
Do multiple attempts (up to 6) of antitachycardia pacing (ATP) improve termination rates of spontaneously occurring VTs in ICD recipients?
Observational (n=24)
Do multiple attempts (up to 6) of antitachycardia pacing (ATP) improve termination rates of spontaneously occurring VTs in ICD recipients?
Answers are grounded in this paper and cite their sources.
The first ATP sequence is the most effective for terminating spontaneous VTs, but up to 6 attempts may be useful in a minority of patients with hemodynamically stable VTs, with a low risk of acceleration in later attempts.
The purpose of this study was to determine the termination and acceleration rates for 1 to 6 attempts of antitachycardia pacing [ATP] delivered by ICD in order to terminate spontaneously occurring VTs. Twenty-four ICD recipients with active ATP programs, including a maximum of six ATP sequences and spontaneously occurring VTs during follow-up, were investigated. During a mean follow-up of 42 +/- 15 months (range, 17-63 months) 413 spontaneous VT episodes (17 +/- 14; range, 1-49 per patient) resulting in appropriate ATP delivery by the ICD occurred. ATP successfully terminated 328 episodes (80%) with a mean number of 1.6 +/- 1.1 pacing sequences. Eighty episodes (19%) were accelerated by ATP and 5 (1%) were unresponsive to ATP. The ATP success decreased until the third ATP sequence (59%-->31%-->24%), but increased again in the fourth to sixth attempt (46%-->46%-->29%). The acceleration rate increased from sequence one to sequence three (8%-->13%-->28%), but decreased significantly in further ATP attempts (19%-->0%-->0%). The mean time delays until redetection or termination after 4, 5, and 6 attempts of ATP were 22 +/- 5 seconds, 37 +/- 2 seconds, and 41 +/- 9 seconds, respectively. Nine patients (37%) used > or = 3 ATP attempts during follow-up and all of them had a therapeutic benefit from it. Five out of 13 VTs (38%) treated with > or = 4 attempts could ultimately be terminated by ATP. The results of this study demonstrate that the first ATP sequence is the most effective and that > 4 ATP attempts may be useful in a minority of patients. There seems to be a low risk of VT acceleration by the fourth to sixth ATP sequence. Because of the associated time delay, a high number of ATP attempts should only be programmed in patients with hemodynamically well-tolerated stable VTs.
No takes yet. Share an insight, caveat, or question.
Fries et al. (1997) conducted an observational in Spontaneously occurring ventricular tachycardias in ICD recipients (n=24). Antitachycardia pacing (ATP) was evaluated on Termination and acceleration rates for 1 to 6 attempts of antitachycardia pacing. Antitachycardia pacing successfully terminated 80% of spontaneous ventricular tachycardia episodes, with the first sequence being the most effective (59% success).