Key result
Abdominal ICD pocket erosion leads to generator migration into the peritoneal cavity.
Why the study?
As ICD therapy use increases, the incidence of recognized and new complications such as pocket erosion and pulse generator migration may increase.
Key points are not available for this paper at this time.
Population
Two patients with ICD therapy
Design
Case report
Answers are grounded in this paper and cite their sources.
Alerts clinicians to rare ICD pocket erosion with peritoneal migration; leaves open incidence, risk factors, and prevention strategies.
ICD therapy for life-threatening arrhythmias is well established. As more patients are treated, the incidence of recognized and new complications may increase. We report ICD pocket erosion and migration of the pulse generator into the peritoneal cavity in two patients.
No takes yet. Share an insight, caveat, or question.
CHITKARA et al. (1996) studied this question. Implantable cardioverter-defibrillator (ICD) pocket erosion and migration of the pulse generator into the peritoneal cavity were observed as complications in two patients.