Answers are grounded in this paper and cite their sources.
Pressure fluctuations in applications such as underground transmission tunnels and high-altitude regions can introduce significant errors in trace gas concentration measurements. This study investigates CO and CO 2 detection under varying pressure conditions (70–103 kPa) using Cavity Ring-Down Spectroscopy (CRDS). A pressure-controlled multi-component CRDS system was developed, and the influence of pressure on the absorption spectra of CO and CO 2 was systematically analyzed. Results show that the full width at half maximum (FWHM) of both spectral lines increases linearly with pressure. Notably, CO exhibits a much larger center wavenumber shift than CO 2 , consistent with its stronger pressure-shift coefficient in the targeted spectral range. The intrinsic loss of the empty cavity is also pressure-dependent, decreasing as pressure drops. To mitigate these effects, a comprehensive pressure compensation algorithm integrating line-shape function modification and cavity baseline calibration was proposed. Experimental validation demonstrates that at 70 kPa, the maximum retrieval errors for CO and CO 2 are reduced from 9.28% and 7.51% to within ±0.5%, respectively. This method significantly improves the accuracy and reliability of CRDS measurements under pressure-varying environments such as transmission tunnels and high-altitude regions.
No takes yet. Share an insight, caveat, or question.
Li et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: