Key result
NexGen bare metal stent yields ~13% 6-month target vessel revascularization rate after PCI.
Why the study?
Despite the dominance of drug-eluting stents, there remains a clinical niche for bare metal stents, necessitating evaluation of the new NexGen cobalt-chromium stent in real-life patients.
Observational (n=383)
Open-label
Yes
Real-world registry data support NexGen BMS feasibility in complex PCI; leaves open randomized comparisons versus contemporary DES.
INTRODUCTION: Despite the dominance of drug-eluting stents in modern interventional cardiology, there is still a niche for bare metal stents. AIM: The aim of the Polish NexGen registry was to evaluate the safety and efficacy of a new generation cobalt-chromium NexGen stent in a real life patient population. MATERIAL AND METHODS: A prospective multi-center registry was conducted in five clinical sites of American Heart of Poland. Three hundred and eighty-three patients who underwent percutaneous coronary intervention (PCI) with NexGen stent implantation were included. Clinical follow-up was performed at 1, 6 and 12 months. Additionally, a group of 42 randomly selected patients underwent control angiography at 6 months (10.96% of study population). The primary endpoint was occurrence of target vessel revascularization (TVR) at 6-month follow-up. Angiographic endpoints included rates of binary restenosis and late lumen loss at 6-month follow-up based on QCA analysis. Multivessel disease was present in more than 70% of patients, and 52.4% of lesions were complex. The main indications for angiography were non-ST elevation acute coronary syndromes (54.8%) and ST elevation myocardial infarction (34.99%). RESULTS: At 6-month follow-up 47 (12.7%) patients reached the primary endpoint of TVR. The composite of major acute cardiac event rates at 30-day and 6- and 12-month follow-up was 6.01% (n = 23), 18.5% (n = 69) and 25.21% (n = 92) respectively. Control angiography performed after 6 months showed in-stent late loss of 0.66 ±0.71 mm and a binary restenosis rate of 16.7%. CONCLUSIONS: Our study showed that PCI with the NexGen stent is safe and effective at 6- and 12-month follow-up. Angiographic results showed a satisfactory restenosis rate and low late lumen loss.
No takes yet. Share an insight, caveat, or question.
Milewski et al. (2016) conducted an observational in Coronary artery disease requiring percutaneous coronary intervention (n=383). NexGen cobalt-chromium bare metal stent was evaluated on Target vessel revascularization (TVR) at 6-month follow-up. The NexGen bare metal stent resulted in a 12.7% rate of target vessel revascularization at 6-month follow-up in a real-world population undergoing percutaneous coronary intervention.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: