Key result
Electronic stethoscope 30-second heart rate measurement shows strong agreement with ECG in healthy cats.
Why the study?
Traditional stethoscopes may not always accurately assess heart rate due to ambient noise interference, while electronic stethoscopes provide amplified signals for more accurate heart sound identification.
Does the 3M Littmann 3200 electronic stethoscope accurately measure heart rate compared to electrocardiography in healthy cats?
Cross-Sectional (n=34)
No
Does the 3M Littmann 3200 electronic stethoscope accurately measure heart rate compared to electrocardiography in healthy cats?
A 30-second measurement using the 3M Littmann 3200 electronic stethoscope provides an accurate and reliable heart rate estimate comparable to ECG in healthy cats.
Answers are grounded in this paper and cite their sources.
May support 30-second electronic stethoscope heart rate assessment in healthy cats; leaves open validation in clinical or diseased populations.
Cardiac auscultation is an indispensable method in the examination of cats. With a traditional stethoscope, heart murmurs, gallop rhythms, and heart rate (HR) may not always be accurately assessed because of ambient noise interference. However, electronic stethoscopes (ESs) provide more uniform and amplified acoustic signals, allowing more accurate identification of physiological and pathological heart sounds during auscultation. The aim of this study was to determine the agreement, differences, and correlation between HR automatically measured using a 3M Littmann 3200 ES over 15-, 30-, and 60-second intervals and HR obtained by electrocardiography (ECG) in 34 healthy cats of various breeds, ages, and sexes. To this end, intraclass correlation coefficient (ICC), Bland–Altman, and Spearman’s correlation analyses were performed. ICC analysis showed moderate agreement between ECG and ES at the 15- and 60-second measurement intervals (r = 0.661, P < 0.001; r = 0.700, P < 0.001, respectively), whereas the 30-second interval demonstrated higher agreement (r = 0.812, P < 0.001). According to Bland-Altman analysis, 30-second HR measurements by ES showed the closest agreement with ECG (bias = 5.03; 95% limits of agreement = -34.34 to 44.40), compared to 15 seconds (bias = 9.41; -48.34 to 67.16) and 60 seconds (bias = 9.06; -41.88 to 59.99). In conclusion, this study demonstrates that 30-second measurements obtained with the 3M Littmann 3200 ES provide heart rate measurements comparable to those obtained by ECG in healthy cats.
No takes yet. Share an insight, caveat, or question.
Ömer Aydın (2026) conducted a cross-sectional in Healthy cats (n=34). 3M Littmann 3200 electronic stethoscope vs. Electrocardiography (ECG) was evaluated on Agreement between 30-second electronic stethoscope heart rate measurement and ECG (Intraclass correlation coefficient) (ICC r = 0.812, p=< 0.001). A 30-second heart rate measurement using the 3M Littmann 3200 electronic stethoscope showed strong agreement with ECG in healthy cats.