Why the study?
No previous study had evaluated the accuracy of a six-lead smartphone-based electrocardiographic device in dogs.
Does a six-lead smartphone-based ECG device provide comparable accuracy to a standard six-lead ECG for assessing heart rate, rhythm, and ECG variables in dogs?
Does a six-lead smartphone-based ECG device provide comparable accuracy to a standard six-lead ECG for assessing heart rate, rhythm, and ECG variables in dogs?
A six-lead smartphone-based ECG device is comparable to standard ECG for heart rate and rhythm assessment in dogs, offering a clinically acceptable alternative for veterinary cardiology.
May support point-of-care rhythm screening in dogs; leaves open need for broader validation across breeds and conditions.
BACKGROUND: No previous study has evaluated the accuracy of a six-lead smartphone-based electrocardiographic (s-ECG) device in dogs. METHODS: This was a prospective multicentre study. Patients referred for cardiologic consultation were selected. In each patient, a 30-second electrocardiogram was simultaneously acquired with a novel six-lead s-ECG and a standard six-lead ECG machine (st-ECG). A board-certified cardiologist evaluated each recording. Nineteen ECG variables, including heart rate and rhythm, as well as quantitative and qualitative features of waves, segments and intervals, were analysed. Agreement between s-ECG and st-ECG was evaluated using Cohen's kappa coefficient and the Bland-Altman test. RESULTS: Seventy-five dogs were enrolled, and 140 ECG tracings were analysed. There was perfect agreement between the two methodologies for heart rate and rhythm classification, both in dogs with sinus rhythm and those with pathological rhythms. Although some disagreement was found when comparing measurements of quantitative variables obtained with the s-ECG and the st-ECG, none of the differences was of clinical relevance. LIMITATIONS: The sample size was limited, and the interobserver variability was not analysed. CONCLUSION: The six-lead s-ECG studied herein is comparable to the st-ECG for heart rate and rhythm assessment, and seems clinically acceptable for the interpretation of waves, segments and intervals in dogs.
No takes yet. Share an insight, caveat, or question.
Romito et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: