Answers are grounded in this paper and cite their sources.
Braided stents were widely used to treat cerebral aneurysms and computational fluid dynamics (CFD) was used to evaluate the therapeutic effects. But the aneurysm-artery geometry used in CFD were usually undeformed which is inconsistent with clinical findings. Our team developed a finite element modeling workflow to simulate implantation of braided stents in patient-specific aneurysm-artery model. An LVIS-based braided stent was deployed into an aneurysm-artery model. The stent fully expanded, causing obvious deformation on the aneurysm-artery model. The workflow which we developed could provide reasonable deformed geometries of aneurysm-artery and braided stent for CFD computation and possibly assist surgical planning.
No takes yet. Share an insight, caveat, or question.
Cai et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: