Key result
Non-complete tissue proximity indication yields ~100% sensitivity for identifying acute PV isolation gaps in PFA.
Why the study?
The spatial distribution of non-complete tissue proximity indicator (TPI) during pulsed field ablation with a variable-loop circular catheter and its relationship to acute isolation gaps had not been systematically evaluated with side-specific analysis.
Does segment-level tissue proximity indicator (TPI) assessment during pulsed field ablation with a variable-loop circular catheter predict acute isolation gaps in patients with atrial fibrillation?
Observational (n=14)
No
Does segment-level tissue proximity indicator (TPI) assessment during pulsed field ablation with a variable-loop circular catheter predict acute isolation gaps in patients with atrial fibrillation?
Non-complete tissue proximity indication during variable-loop pulsed field ablation shows 100% sensitivity and 100% negative predictive value for acute pulmonary vein isolation gaps, enabling targeted additional ablation.
Non-complete TPI may flag acute PVI gaps during variable-loop PFA; leaves open whether TPI-guided re-ablation improves durable isolation.
Background/Purpose Pulsed field ablation (PFA) using a variable-loop circular catheter is an emerging modality for pulmonary vein isolation (PVI). The tissue proximity indicator (TPI) provides real-time feedback on electrode-tissue contact; however, the spatial distribution of non-complete TPI and its relationship to acute isolation gaps have not been systematically evaluated with side-specific analysis. We performed the first segment-level spatial analysis of TPI patterns using a novel eight-segment classification, separately for left and right PVs. Methods In this prospective single-centre study, 14 consecutive patients with atrial fibrillation (AF) underwent PVI using a variable-loop circular PFA catheter. For each ipsilateral PV pair, two ablations were delivered within each PV and six at the antrum encompassing both veins, totalling ten ablations per side. Post-ablation TPI was assessed using the TPI filter on the electroanatomical mapping system. Each PV ostium was divided into eight segments: superior, anterior-superior, anterior-carina, anterior-inferior, inferior, posterior-inferior, posterior-carina, and posterior-superior. TPI was graded as complete, partial, or poor. Voltage mapping identified acute isolation gaps. Left PVs (LPVs; 112 segments) and right PVs (RPVs; 112 segments) were analysed separately. Results Mean age was 69±10 years, 50% were female, and 50% had persistent AF. Procedure and fluoroscopy times were 58±9 and 14±3 min, respectively. Overall complete TPI rates were similar between LPVs and RPVs (94.6% vs 93.8%). However, vulnerability patterns differed markedly: in RPVs, the anterior-superior segment was most frequently non-complete (28.6%), harbouring 3 of 5 RPV gap sites. In LPVs, non-complete TPI was distributed across the superior (14.3%) and anterior-inferior (14.3%) segments. The posterior-carina was vulnerable bilaterally (LPV 7.1%; RPV 14.3%). The inferior and posterior-inferior segments achieved 100% complete TPI in both LPVs and RPVs, with no gaps at these sites. Acute gaps were detected in 4/14 LPVs (5 sites) and 5/14 RPVs (5 sites). Critically, all 10 gap sites across both sides exhibited non-complete TPI (sensitivity 100%), and no gaps occurred where TPI was complete (NPV 100%; p<0.0001 for both). No procedural complications occurred. Conclusions Eight-segment spatial analysis revealed distinct side-specific vulnerability patterns during PFA. The anterior-superior segment of RPVs and the superior and anterior-inferior segments of LPVs were most susceptible to incomplete tissue proximity. Non-complete TPI demonstrated perfect concordance with acute isolation gaps bilaterally, suggesting that segment-level TPI assessment can reliably predict incomplete isolation and guide targeted additional ablation.
No takes yet. Share an insight, caveat, or question.
Yoshizawa et al. (2026) conducted an observational in Atrial fibrillation (n=14). Non-complete tissue proximity indicator (TPI) during pulsed field ablation vs. Complete tissue proximity indicator (TPI) was evaluated on Concordance of non-complete tissue proximity indicator (TPI) with acute isolation gaps (p=<0.0001). Non-complete tissue proximity indication during variable-loop circular catheter PFA showed 100% sensitivity and 100% NPV for acute isolation gaps (p<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: