Key result
Novel CSP morphology monitor algorithm detects capture changes with ~99% negative predictive value.
Why the study?
The need for 12-lead ECGs during follow-up of pacemaker patients with conduction system pacing presents a clinical workflow challenge, motivating the evaluation of a new algorithm that automatically monitors ventricular electrogram morphology.
Does the novel pacemaker-based CSP morphology monitor algorithm accurately detect changes in conduction system pacing capture compared to expert 12-lead ECG adjudication in pacemaker patients with left bundle branch area pacing?
Observational (n=38)
Does the novel pacemaker-based CSP morphology monitor algorithm accurately detect changes in conduction system pacing capture compared to expert 12-lead ECG adjudication in pacemaker patients with left bundle branch area pacing?
Answers are grounded in this paper and cite their sources.
A novel pacemaker-based electrogram morphology monitor algorithm shows high specificity (>92%) and negative predictive value (>97%) for confirming ongoing conduction system pacing capture, which may reduce the need for routine in-office 12-lead ECG follow-ups.
Introduction The need for 12-lead ECGs during follow-up of pacemaker patients with conduction system pacing (CSP) presents a clinical workflow challenge. CSP morphology monitor is a new algorithm designed to automatically monitor ventricular electrogram (EGM) morphology, aiding clinicians in confirming and maintaining effective CSP therapy. Purpose To evaluate the performance of the CSP morphology monitor algorithm in a clinical test dataset, with specificity and negative predictive value (NPV) as the primary end point. Methods To assess CSP capture, changes in far-field EGM (RV Tip-to-Can) were calculated using two metrics (Figure 1): Wavelet Match Score (0–100%), which quantifies similarity between a stored in-office template beat (e.g., at 5 V) and subsequent test beats using a Wavelet transform; and EGM-derived Left Ventricular Activation Time (LVAT), measured from the ventricular pacing marker to the center of the R-wave area, correlating with ECG LVAT. Prospective data were collected in pacemaker patients with left bundle branch area pacing using the Model 3830 lead. Device EGMs and 12-lead ECGs were acquired during pacing threshold tests at follow-up (unipolar/bipolar pacing; supine/sitting). Performance was assessed by comparing beat-by-beat algorithm outputs with expert adjudication using generalized estimating equations (thresholds: 70% Match Score, ±12% EGM LVAT Ratio). Results A total of 38 patients were included (mean age: 70 years, range 27–89; 45% male; follow-up visits 1-9 months post-implant). EGM templates and EGM LVAT during non-selective left bundle branch pacing (NS-LBBP) at 5 V were successfully created in all patients. The algorithm was evaluated with 11,346 observations for Wavelet Match Score and 11,331 for EGM LVAT. Changes in EGM morphology mainly included transitions from NS-LBBP to left ventricular septal pacing (LVSP) or selective LBBP (S-LBBP). Beats with loss of pacing capture were excluded. The algorithm demonstrated a specificity of 92.8% (95% confidence interval [CI]: 89.1–95.3%), and NPV of 98.5% (95% CI: 96.9–99.3%) for Wavelet Match Score; and a specificity of 97.1% (95% CI: 93.3–98.8%) and NPV of 97.2% (95% CI: 93.2–98.9%) for EGM LVAT in detecting changes in CSP capture. Of the 12 patients with transitions from NS-LBBP to LVSP, 10 (83%) were correctly identified using either the Wavelet Match Score or the EGM LVAT algorithm. The 2 patients who transitioned from NS-LBBP to S-LBBP were detected by the Wavelet Match Score algorithm, while EGM LVAT changes remained within 12%. Conclusion The CSP morphology monitor showed high specificity and negative predictive value for both Wavelet Match Score and EGM-derived LVAT. When Match Score is ≥ 70% or LVAT changes remain within 12%, a change in NS-LBBP capture is unlikely. This algorithm can detect clinically relevant EGM morphology changes and could help reduce unnecessary in-office follow-up visits for future pacemaker patients.
No takes yet. Share an insight, caveat, or question.
Vijayaraman et al. (2026) conducted an observational in Conduction system pacing (left bundle branch area pacing) (n=38). CSP morphology monitor algorithm (Wavelet Match Score and EGM-derived LVAT) vs. Expert adjudication of 12-lead ECGs and device EGMs was evaluated on Specificity and negative predictive value (NPV) for detecting changes in CSP capture (Wavelet Match Score specificity 92.8%, NPV 98.5%; EGM LVAT specificity 97.1%, NPV 97.2%, 95% CI 89.1-95.3). The CSP morphology monitor algorithm achieved 92.8% specificity and 98.5% NPV via Wavelet Match Score, and 97.1% specificity and 97.2% NPV via EGM LVAT for detecting CSP capture changes.