Key result
Subcutaneous ICDs match transvenous ICDs for 8-year appropriate therapy rates and first-shock efficacy.
Why the study?
Although subcutaneous and transvenous ICDs showed no differences in appropriate therapy rates or first-shock efficacy during 4.1 years of follow-up in the PRAETORIAN trial, data on long-term incidence and efficacy of appropriate therapy were lacking.
Does a subcutaneous implantable cardioverter-defibrillator provide comparable long-term incidence and efficacy of appropriate therapy compared to a transvenous implantable cardioverter-defibrillator in patients with a class I or IIa ICD indication without pacing needs?
RCT (n=849)
Randomized (modified intention-to-treat analysis)
Yes
Does a subcutaneous implantable cardioverter-defibrillator provide comparable long-term incidence and efficacy of appropriate therapy compared to a transvenous implantable cardioverter-defibrillator in patients with a class I or IIa ICD indication without pacing needs?
Over a median 7.3 years of follow-up, subcutaneous and transvenous ICDs demonstrated comparable rates of appropriate therapy and first-shock efficacy, with a continuous linear increase in new patients receiving appropriate therapy supporting the long-term benefit of ICDs.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“This is the longest follow-up of patients randomized to subcutaneous or transvenous ICDs, offering unique insight into their long-term safety and efficacy.”
“We therefore felt that it was desirable to follow up this group of patients for a longer period”
Supports long-term S-ICD/TV-ICD equivalence for appropriate therapy and shock efficacy; extends 4-year PRAETORIAN findings to median 7.3 years.
Background Previous secondary analyses of the PRAETORIAN trial showed no differences in all appropriate therapy (shocks and antitachycardia pacing (ATP)) rates or first-shock efficacy between subcutaneous (S-ICD) and transvenous ICDs (TV-ICD) during 4.1 years of follow-up. However, ICD therapy is intended to provide lifelong protection, and data on the incidence and efficacy of appropriate therapy during long-term follow-up are lacking. The PRAETORIAN-XL trial extends follow-up by an additional 4 years to evaluate long-term appropriate therapy rates and first-shock efficacy. We hypothesize that S-ICD and TV-ICD demonstrate comparable performance during extended follow-up of more than 8 years. Purpose To compare the long-term incidence and efficacy of appropriate therapy in patients with S-ICD and TV-ICD, a secondary analysis of the PRAETORIAN-XL trial. Methods This is a secondary analysis of the PRAETORIAN-XL trial. From 39 centers in Europe and the United States, patients with a class I or IIa indication for implantable cardioverter-defibrillator (ICD) without the need for cardiac pacing were enrolled and randomized between March 2011 and January 2017. Patients were followed for an additional 4 years after the follow-up period of the PRAETORIAN trial. Appropriate therapy was defined as ATP or ICD shock for ventricular fibrillation (VF) or ventricular tachycardia (VT), and electrical storm was defined as 3 or more VF or VT episodes within 24 hours. Only appropriate therapies adjudicated with electrograms were analyzed, and analyses were performed according to a modified intention-to-treat principle. Results A total of 849 patients were randomized to the S-ICD group (n=426) and the TV-ICD group (n=423). During a median 7.3 (interquartile: 3.8-8.7) years follow-up, there was no difference in appropriate therapy rate between the S-ICD group and the TV-ICD group (116 and 103 patients, respectively; 8-year Kaplan-Meier estimated cumulative incidence: 31.7% vs. 29.8%; p=0.293) (Figure 1, 2). During follow-up, there is a linear annual increase in new patients receiving appropriate therapy (S-ICD: 2.6-8.6 per 100 patient-years, and TV-ICD: 3.1-6.8 per 100 patient-years). The first shock efficacy was similar between the S-ICD groups and the TV-ICD group (92.4% vs. 92.7%; p=0.888). The first ATP terminated 54.4% of monomorphic VTs (160/294), however, 9.9% of VT were accelerated and 0.7% of VT were converted to VF after ATP delivery (Figure 1). In total, 22 electrical storms occurred in 16 patients in the S-ICD group, and 24 electrical storms occurred in 20 patients in the TV-ICD group (p=0.482). Conclusions During long-term follow-up, there were no differences in the rate of appropriate therapy and first shock efficacy between S-ICD and TV-ICD. There is a linear increase in new patients receiving appropriate therapy over the years, emphasizing the long-term benefit of ICD therapy with the current guidelines.Figure 1.The details of the therapiesFigure 2.
No takes yet. Share an insight, caveat, or question.
Matsuda et al. (2026) conducted an RCT in Indication for implantable cardioverter-defibrillator (ICD) without need for cardiac pacing (n=849). Subcutaneous implantable cardioverter-defibrillator (S-ICD) vs. Transvenous implantable cardioverter-defibrillator (TV-ICD) was evaluated on Appropriate therapy (antitachycardia pacing or ICD shock for ventricular fibrillation or ventricular tachycardia) at 8 years (p=0.293). Subcutaneous ICDs showed no difference in 8-year appropriate therapy rates (31.7% vs. 29.8%; p=0.293) or first-shock efficacy (92.4% vs. 92.7%; p=0.888) compared with transvenous ICDs.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: