Key result
H2-CBP wrist device tracks brachial cuff BP with ~3 mmHg systolic bias across rest and stress.
Why the study?
Wrist-based cuffless BP monitoring offers potential advantages for longitudinal and ambulatory assessment, but signal instability and measurement variability under physiologically dynamic conditions have limited its use. Agreement between a novel wrist-worn dual-sensor applanation tonometry device (H2-CBP) and reference brachial BP across rest, stress, and recovery required evaluation.
Does a wrist-worn dual-sensor applanation tonometry device (H2-CBP) demonstrate accurate agreement with reference brachial cuff-based blood pressure measurements across rest, exercise, and recovery conditions in adult participants?
Observational (n=25)
Does a wrist-worn dual-sensor applanation tonometry device (H2-CBP) demonstrate accurate agreement with reference brachial cuff-based blood pressure measurements across rest, exercise, and recovery conditions in adult participants?
The H2-CBP wrist-worn dual-sensor applanation tonometry device demonstrated high reproducibility and strong agreement with reference brachial cuff blood pressure across rest, exercise, and recovery conditions, showing promise for longitudinal blood pressure monitoring.
Answers are grounded in this paper and cite their sources.
May support wearable BP research applications; leaves open clinical adoption pending larger validation studies.
Background Wrist-based cuffless blood pressure (BP) monitoring offers potential advantages for longitudinal and ambulatory assessment but has been limited by signal instability and measurement variability, particularly under physiologically dynamic conditions. We evaluated the agreement between a novel wrist-worn blood pressure monitoring device incorporating dual-sensor applanation tonometry (H2-CBP) and reference brachial blood pressure across rest, stress, and recovery conditions. Methods In this prospective validation study, 25 adult participants underwent paired blood pressure measurements using H2-CBP and a validated automated brachial cuff-based device. Measurements were obtained at rest, during a standardized seated lower-limb flexion–extension exercise protocol (5 minutes), and 1 hour after exercise. Agreement was assessed using Bland–Altman analysis, root mean square error (RMSE), and intraclass correlation coefficients (ICC). Linear association was evaluated using Pearson correlation and linear regression analyses. Subgroup analyses were performed by sex and body mass index (BMI). Results For systolic blood pressure (SBP), the overall mean bias was −3.37 mmHg, with limits of agreement (LOA) from −10.60 to 3.87 mmHg and RMSE of 4.68-5.31 mmHg across conditions. For diastolic blood pressure (DBP), mean bias was −1.64 mmHg, with LOA from −9.97 to 6.68 mmHg and RMSE of 4.01-4.54 mmHg. ICC were consistently high for both SBP and DBP (≥0.94 across all conditions). Strong linear associations were observed between H2-CBP and reference measurements (SBP r=0.972–0.978; DBP r=0.891–0.921; all P<0.001). Agreement metrics were preserved across sex and BMI subgroups. Conclusions H2-CBP demonstrated robust agreement with reference brachial BP across rest, exercise, and recovery conditions, with low bias, modest error, and high reproducibility. The dual-sensor applanation tonometry design appears to mitigate key limitations of prior single-sensor wrist-based devices. Although not intended to replace diagnostic cuff-based measurements, H2-CBP shows promise as a tool for longitudinal and context-aware blood pressure monitoring in clinical and research settings.
No takes yet. Share an insight, caveat, or question.
Jung et al. (2026) conducted an observational in Blood pressure monitoring across rest, exercise, and recovery conditions (n=25). Wrist-worn dual-sensor applanation tonometry device (H2-CBP) vs. Validated automated brachial cuff-based blood pressure device was evaluated on Agreement in systolic and diastolic blood pressure (Bland-Altman mean bias, limits of agreement, RMSE, and ICC) (MD -3.37 mmHg (SBP); MD -1.64 mmHg (DBP), p=<0.001). The H2-CBP wrist device showed strong agreement with brachial cuff BP across rest and stress.