Key result
Direct iASD closure through MitraClip steerable guide catheter achieves 0% complications with rapid occlusion.
Why the study?
Is direct intra-procedural closure of an iatrogenic atrial septal defect through the MitraClip steerable guide catheter safe and feasible?
Observational (n=11)
No
Is direct intra-procedural closure of an iatrogenic atrial septal defect through the MitraClip steerable guide catheter safe and feasible?
Direct intra-procedural closure of iatrogenic atrial septal defects through the MitraClip steerable guide catheter is a safe technique that maintains continuous left atrial access.
Answers are grounded in this paper and cite their sources.
May facilitate iASD closure without losing LA access during MitraClip; leaves open need for prospective safety and efficacy data.
OBJECTIVES: To introduce a novel method of direct iatrogenic atrial septal defect (iASD) closure through the MitraClip steerable guide catheter (SGC). BACKGROUND: MitraClip implantation requires transseptal puncture and the creation of an iASD. There are relatively rare instances, such as hemodynamically significant shunting or concerns for embolus, where iASD must be closed during index procedure. In these instances, it may be beneficial to not give up access to left atrium. METHODS: We retrospectively reviewed all iASD closures during MitraClip implantation at our institution from 2015 to 2020. Cases where an ASD occluder was deployed directly through SGC were included. RESULTS: Eleven patients had immediate iASD closure through the SGC. Indications for using this method included concern for paradoxical embolus, large defect size and/or significant shunting. Closure device sizes ranged from 8 to 22 mm. Mean time from removal of clip delivery system to occlusion of iASD was 14.6 minutes. There were no procedural complications related to iASD closure using this method. CONCLUSION: Closure of iASD intra-procedurally directly through transseptal guide sheath via the method described was safe and allowed for continuous left atrium access.
No takes yet. Share an insight, caveat, or question.
Bernstein et al. (2022) conducted an observational in Iatrogenic atrial septal defect (iASD) during MitraClip implantation (n=11). Direct iatrogenic atrial septal defect (iASD) closure through the MitraClip steerable guide catheter (SGC) using an ASD occluder (8 to 22 mm) was evaluated on Procedural complications related to iASD closure. Direct intra-procedural iASD closure through the MitraClip steerable guide catheter resulted in 0% procedural complications and a mean occlusion time of 14.6 minutes across 11 patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: