Novel textile electrodes and Kalman filtering boost wearable ECG signal-to-noise ratio by ~9 dB.
Why the study?
Prior ECG denoising studies operated offline, leaving a need for a real-time extended Kalman filtering framework to extract motion and electromyography artifacts in wearable systems.
Answers are grounded in this paper and cite their sources.
Real-time wearable ECG artifact removal is feasible; hypothesis-generating and requires prospective validation before clinical adoption.
In this paper, a prototype of wearable and wireless electrocardiography (ECG) monitoring system is developed and implemented on DSP and PDA. We present a real-time extended Kalman filtering framework for extracting motion and electromyography (EMG) artifacts from a single-channel ECG in wearable systems as different from other offline studies. Realized prototype is a good example for the usage of the Kalman filter in biomedical real-time system. The average SNR advancement of 9.1430 dB was achieved for denoising, which is average 1 dB more than the other methods such as MABWT, EKF2 by using MIT-BIH database. Additionally, the usability and performances of conductive textile electrodes were evaluated with disposable Ag – AgCl electrodes by using daily activities. A novel textile electrode gave approximately 25.23% better results compared to Ag – AgCl electrodes. Also, UDP, TCP and Web Socket communication protocols have been tested. UDP has been the fastest method for the ECG signal transferring from the patient to the doctor. At the same time, a method is proposed for direct access to the patient by the doctor. The results illustrate that this type of system will submit highly ergonomic solutions among biomedical device technologies. In addition, the usage of such kinds of systems is foreseen for requiring long-term follow-up and disorders.
No takes yet. Share an insight, caveat, or question.
Özkaraca et al. (2014) studied this question. Real-time extended Kalman filter with conductive textile electrodes in a wearable ECG system vs. MABWT, EKF2, and disposable Ag-AgCl electrodes was evaluated on ECG signal denoising (SNR advancement) and electrode performance. A wearable ECG system with real-time extended Kalman filtering achieved an average SNR advancement of 9.1430 dB, while novel textile electrodes performed ~25.23% better than Ag-AgCl electrodes.
Synapse has enriched 10 related papers. Consider them for comparative context: