Key result
Remote monitoring detects ICD lead fracture via electrical interference despite normal lead impedances.
Population
A patient with an ICD lead fracture at the third device replacement
Answers are grounded in this paper and cite their sources.
Electrical interference on remote monitoring may flag ICD lead fracture despite normal impedances; hypothesis-generating for enhanced surveillance protocols.
We described the fracture of an ICD lead in a patient at the third device replacement.Through the remote monitoring electric interference was detected on the defibrillator lead despite the pacing and shock impedances during the leadlifetime remained stable and in a normal range. X-ray confirmed the lead failure and patient underwent to a lead replacement.
No takes yet. Share an insight, caveat, or question.
Scicchitano et al. (2023) studied this question. Remote monitoring detected electrical interference indicative of an ICD lead fracture despite stable, normal pacing and shock impedances.