Key result
Pulsed field ablation in CIED patients is linked to a ~4% incidence of unexpected VF.
Why the study?
What is the procedural incidence of unexpected ventricular fibrillation during pulsed field ablation in patients with cardiac implantable electronic devices?
Cohort (n=107)
No
What is the procedural incidence of unexpected ventricular fibrillation during pulsed field ablation in patients with cardiac implantable electronic devices?
Unexpected ventricular fibrillation occurred in 3.7% of pulsed field ablation procedures in patients with CIEDs, underscoring the need for systematic periprocedural device management and immediate availability of external defibrillation.
Answers are grounded in this paper and cite their sources.
May warrant immediate defibrillation availability during pulsed field ablation in CIED patients; hypothesis-generating for prospective safety studies.
BACKGROUND: Arrhythmia induction during catheter ablation in patients with cardiac implantable electronic devices (CIEDs) has been reported, but the risk associated with pulsed field ablation (PFA) remains poorly characterized. OBJECTIVE: To estimate the procedural incidence of unexpected ventricular fibrillation (VF) during PFA in patients with CIEDs and describe associated clinical and device findings. METHODS: We performed a retrospective single-center cohort study of consecutive PFA procedures in patients with CIEDs. All procedures were reviewed for unexpected VF. RESULTS: Among 107 consecutive procedures using PFA energy in patients with CIEDs, 83 were atrial and 24 were ventricular ablations. Unexpected VF occurred in 4 patients (overall incidence 3.7%, 95% CI 1.0% - 9.3%): 1 during left atrial PFA for pulmonary vein isolation and 3 during left ventricular PFA for ventricular tachycardia ablations. VF occurred at ablation sites both near and remote from CIED lead tips. Device electrograms demonstrated prominent ablation artifact with abnormal triggered activity and mode switch behavior. Intra-procedural recordings also demonstrated local capture and R on T mechanisms. CONCLUSIONS: Unexpected VF during PFA in patients with CIEDs occurred in 3.7% of procedures. These results complement prior adverse-event reports and support systematic pre-, intra-, and post-procedure CIED management, with external defibrillation immediately available, whenever PFA or dual-energy ablation is performed in patients with cardiac devices.
No takes yet. Share an insight, caveat, or question.
Kong et al. (2026) conducted a cohort in Patients with cardiac implantable electronic devices (CIEDs) undergoing pulsed field ablation (PFA) (n=107). Pulsed field ablation (PFA) with dual-energy focal catheters was evaluated on Procedural incidence of unexpected ventricular fibrillation (VF) (Incidence 3.7%, 95% CI 1.0%-9.3%). Pulsed field ablation in patients with CIEDs was associated with a 3.7% (95% CI 1.0%-9.3%) procedural incidence of unexpected ventricular fibrillation.