Blood pumps usually use elastomer diaphragms that undergo repeated small-strain flexing (deformation). The stresses and strains that develop in the mechanically actuated diaphragms dictate the useful life and fluid-solid interactions of the diaphragm. A knowledge of these stresses and strains is essential for proper design of the pump's elastomer diaphragm. A nonlinear axisymmetric finite-element stress-strain analysis has been carried out on a Hexsyn rubber diaphragm used in a blood pump to illustrate a finite-element modeling technique. A situation involving the application of an axial deformation load on the diaphragm has been considered. In particular, the strains developed in a blood pump diaphragm during pumping have been quantified to illustrate the analysis technique.