Answers are grounded in this paper and cite their sources.
Retaining a lubricating, drug-delivering hydrogel on a load-supporting implant requires coordinated scaffold and interface design. We developed an osteochondral-inspired poly(ether ether ketone) (PEEK)/hydrogel bilayer by continuous 3D printing and surface modification. A dense PEEK base supports a porous upper region containing a poly(vinyl alcohol)/poly(acrylic acid) (PVA/PAA) composite hydrogel. Sulfonation and ultraviolet-assisted acrylic acid (AA) treatment prepare the substrate for hydrogel attachment. At a 4:1 PVA:AA feed mass ratio, the drug-free composite construct (gPCH) exhibited a maximum detachment force per unit width of 304.60 ± 7.20 N m−1, 2.39 times that of PVA on sulfonated PEEK without AA/UV treatment. Its friction coefficient was 0.073 ± 0.005 during a 10 min screening test under 2 N in 25% fetal bovine serum at room temperature. The diclofenac sodium (DFN)-loaded counterpart (gPCHD) released 73.92 ± 3.24% of DFN over 72 h in PBS at 37 °C. In lipopolysaccharide-stimulated rabbit bone marrow-derived mesenchymal stem cells, gPCHD showed lower COX-1, COX-2, and IL-6 mRNA expression than composition-matched, drug-free gPCH. Twelve-week rabbit histology (n = 3 per group) provided descriptive observations of tissue contact and a retained hydrogel-bearing layer. This architecture–interface design integrates structural support, interfacial retention, short-duration lubrication, and drug delivery; long-term tribological performance and component-specific in vivo effects remain to be established.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: