A deuterated (EJ315) and a conventional hydrogen based (EJ301) liquid scintillator detectors having same dimension (diameter × length: 2”× 2”), were characterized with quasi-monoenergetic neutrons generated using 7 Li(p,n) 7 Be and 11 B(p,n) 11 C reactions. Scintillator light output was measured in the neutron energy range from 1 to 10 MeV, and its functional dependence on recoil deuteron/proton energy was determined. These functional forms were used in the Geant4 to simulate the response matrices for the two scintillators. These matrices were used to unfold the measured pulse-height distribution, which determined the neutron energy spectrum. These unfolded neutron spectra were in good agreement with those determined using the time-of-flight method. Additionally, a comparative study of pulse shape discrimination and neutron detection efficiency is presented.
No takes yet. Share an insight, caveat, or question.
Dalal et al. (2026) studied this question.