Key result
Aristotle Colossus macrowire enables microcatheter-free aspiration catheter delivery in ~88% of anterior circulation thrombectomies.
Why the study?
Larger macrowires may facilitate microcatheter-free delivery of large-bore aspiration catheters for intracranial endovascular mechanical thrombectomy, motivating an evaluation of the safety and efficacy of the Aristotle Colossus guidewire for this use.
Does the 0.035-inch Aristotle Colossus guidewire facilitate safe and effective microcatheter-free delivery of large-bore aspiration catheters in patients undergoing anterior circulation endovascular mechanical thrombectomy?
Observational (n=122)
Yes
Does the 0.035-inch Aristotle Colossus guidewire facilitate safe and effective microcatheter-free delivery of large-bore aspiration catheters in patients undergoing anterior circulation endovascular mechanical thrombectomy?
The 0.035-inch Aristotle Colossus macrowire safely facilitates microcatheter-free delivery of large-bore aspiration catheters in 87.7% of anterior circulation ischemic stroke thrombectomies, though severe ICA tortuosity and intracranial atherosclerosis reduce success odds.
Answers are grounded in this paper and cite their sources.
May support microcatheter-free aspiration in select anterior circulation cases; leaves open prospective validation before adoption.
BackgroundLarger "macrowires" may facilitate microcatheter-free delivery of large-bore aspiration catheters (ACs) for intracranial endovascular mechanical thrombectomy. We examined the safety and efficacy of the Aristotle Colossus Guidewire for this use.MethodsIn this retrospective analysis of prospective data from 12 institutions, we identified anterior circulation endovascular mechanical thrombectomy cases involving the Aristotle Colossus guidewire over a 12-month period. The primary efficacy endpoint was successful delivery of the AC to the site without the use of adjunct devices. Multivariable logistic regression was performed with clinically relevant covariables.ResultsIn 122 patients (median age 74 years (interquartile range (IQR) 63-82)), most intracranial occlusions were in the M1 segment of the middle cerebral artery (95/122, 78.0%). The primary endpoint of microcatheter-free AC delivery was met in 87.7% (107/122) of patients; 93/122 (76.2%) patients underwent microcatheter-free delivery with successful reperfusion via contact aspiration (modified thrombolysis in cerebral infarction grade ≥ 2B). The median number of passes was 1 (IQR 1-2). Multivariable regression demonstrated that severe internal carotid artery tortuosity or cervical loop presence (odds ratio (OR): 0.09, 95% confidence interval (CI): 0.01-0.54, p = 0.01) and presence of intracranial atherosclerotic disease (OR: 0.15, 95% CI: 0.03-0.83, p = 0.03) were significantly associated with lower odds of successful AC delivery over the guidewire. Intraprocedural vasospasm occurred in 4.9% (6/122) of cases; there were no reports of vessel dissection or perforation.ConclusionsWe demonstrate the favorable performance of the Aristotle Colossus guidewire for anterior circulation ischemic stroke intervention. The macrowire appears to safely facilitate navigation of the neurovasculature by large-bore ACs without the need for microcatheter and/or stentriever.
No takes yet. Share an insight, caveat, or question.
Ares et al. (2025) conducted an observational in Anterior circulation ischemic stroke undergoing endovascular mechanical thrombectomy (n=122). 0.035-inch Aristotle Colossus guidewire for microcatheter-free delivery of large-bore aspiration catheters was evaluated on Successful delivery of the aspiration catheter to the occlusion site without the use of adjunct devices (microcatheter-free delivery). The Aristotle Colossus macrowire achieved microcatheter-free aspiration catheter delivery in 87.7% (107/122) of anterior circulation thrombectomy patients.
Synapse has enriched 10 related papers. Consider them for comparative context: