Key result
Aristotle Colossus 0.035-inch macrowire achieves ~87% TICI 2C/3 recanalization in mechanical thrombectomy.
Why the study?
Catheter-to-microwire step-off and distal navigation remain ongoing challenges during mechanical thrombectomy with large-bore aspiration catheters, and early experience with the new Aristotle 35 Colossus 0.035-inch macrowire to address these challenges had not been reported.
Does the Aristotle 35 Colossus 0.035' macrowire facilitate safe and effective catheter navigation and recanalization without clot traversal in patients undergoing mechanical thrombectomy?
Observational (n=30)
Yes
Does the Aristotle 35 Colossus 0.035' macrowire facilitate safe and effective catheter navigation and recanalization without clot traversal in patients undergoing mechanical thrombectomy?
Answers are grounded in this paper and cite their sources.
Early experience demonstrates that the 0.035' Aristotle Colossus macrowire enables safe triaxial delivery of large-bore aspiration catheters to the clot interface without requiring thrombus traversal, achieving high recanalization rates (87% TICI 2C/3) with zero wire-related perforations.
Introduction Recent literature continues to demonstrate the successful role of large-bore aspiration catheters in thrombus ingestion during mechanical thrombectomy. However catheter-to-microwire step-off and distal navigation are ongoing challenges in thrombectomy. A new to market 0.035’ macrowire (Aristotle 35 Colossus Guidewire, Scientia Vascular, West Vale City, UT) may address such challenges. We report here our early experience in 30 mechanical thrombectomy cases. Materials and Methods We analyzed a prospectively maintained database of the senior authors to identify cases utilizing a 0.035’ macrowire with 0.035’ aspiration catheters for mechanical thrombectomy. Results Thirty consecutive cases were identified. Seventeen (57%) patients were female with an average age of 75.3 ± 2.2 years (range 55–97). Average presenting NIHSS was 13.0 ± 1.7. Thrombus locations included 7% ( n = 2) in the cervical ICA, 47% ( n = 14) in the M1, 43% ( n = 13) in the M2, and 3% ( n = 1) in the P1. An 088’ ID aspiration catheter was navigated to at least the M1 segment in all anterior circulation cases and the basilar in the posterior circulation case. The 0.035’ macrowire was placed proximal to the occlusion in all cases allowing coaxial 035’ and 071’ catheter aspiration passes. TICI 2C/3 was achieved in 87% of cases ( n = 26) and TICI 2B in the remaining cases. There were no wire-related perforations or vessel dissections. Conclusion The Colossus 0.035’ macrowire may offer advantages over its smaller counterparts by reducing ledge effect and the need to cross the thrombus. Further comparative studies against currently available microwires in various anatomies are warranted.
No takes yet. Share an insight, caveat, or question.
Zarrin et al. (2024) conducted an observational in Acute ischemic stroke undergoing mechanical thrombectomy (n=30). Aristotle 35 Colossus 0.035' macrowire was evaluated on Technical success (positioning at the level of the thrombus without traversal) and TICI 2C/3 recanalization. The Aristotle Colossus 0.035' macrowire achieved 100% technical success and 87% TICI 2C/3 recanalization without wire-related perforations or dissections in 30 mechanical thrombectomy cases.