Why the study?
Radiation therapy poses an incompletely characterized risk to CIEDs, and while malfunction mechanisms have been described for transvenous devices, there is a paucity of data for subcutaneous implantable cardioverter-defibrillators (S-ICD).
Case Report (n=1)
No
Pelvic radiation therapy can cause irreversible software memory errors and incessant power-on resets in an out-of-field, shielded S-ICD, which can be safely bridged with a wearable defibrillator until definitive transvenous ICD revision.
Answers are grounded in this paper and cite their sources.
May warrant caution with S-ICDs during pelvic radiation; extends transvenous risks but leaves generalizability open.
The management of patients with cardiac implantable electronic devices (CIEDs) undergoing radiation therapy (RT) requires balancing oncologic necessity with device safety. RT poses a recognized but incompletely characterized risk to CIEDs. For transvenous CIEDs, several mechanisms of malfunction in the setting of radiation have been described, such as semiconductor damage, electromagnetic interference, and neutron scatter, yet there is a paucity of data for subcutaneous implantable cardioverter-defibrillators (S-ICD).
No takes yet. Share an insight, caveat, or question.
Rayfield et al. (2026) conducted a case report in Brugada syndrome with subcutaneous implantable cardioverter-defibrillator (S-ICD) undergoing pelvic radiation therapy for prostate adenocarcinoma (n=1). Pelvic external beam radiation therapy (IMRT) and low dose rate brachytherapy; managed via S-ICD deactivation, wearable cardioverter-defibrillator bridging, and subsequent S-ICD extraction with transvenous ICD implantation was evaluated on S-ICD system malfunction (incessant power-on resets and memory error) and clinical resolution following device revision.