Answers are grounded in this paper and cite their sources.
Structure-switching aptamers are a promising class of molecular recognition elements for continuous biosensing. However, commonly used readout approaches are based on bulk signal from the entire receptor population, and are thus vulnerable to drift due to various extrinsic factors unrelated to target binding. We introduce a time-resolved Förster resonance energy transfer (TR-FRET) sensing strategy that resolves contributions from distinct aptamer conformational states, enabling a robust, self-normalized readout. We label the aptamer with a long-lived europium chelate donor and a Cy5 acceptor and then record the sensitized acceptor emission following pulsed donor excitation. This produces a decay profile whose shape encodes the relative populations of aptamer conformational statesa readout that is insensitive to extrinsic factors that scale signal intensity, such as aptamer concentration or transduction efficiency. We apply this approach to lactate- and dopamine-binding aptamers and demonstrate target-dependent sensing in physiologically relevant buffer conditions. Importantly, the TR-FRET readout remains stable even under aggressive nuclease degradation that reduces fluorescence intensity by over 50%, a condition that renders intensity-based readouts uninterpretable. These results establish TR-FRET decay analysis as a robust, self-normalized readout for structure-switching aptamer sensors.
No takes yet. Share an insight, caveat, or question.
Gopalan et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: