Key result
Impedance drop-guided RF ablation with the TFSE catheter achieves ~93% durable PVI at remapping.
Why the study?
Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation catheter to guide pulmonary vein isolation (PVI) are limited, and conventional impedance drop is hampered by cardiac and respiratory oscillation artifacts.
Does a point-by-point radiofrequency ablation protocol targeting a proportional average impedance drop (%AID) ≥9% using the TactiFlex Sensor Enabled catheter achieve high first-pass isolation and durable pulmonary vein isolation at mandatory remapping in patients with atrial fibrillation?
Observational (n=26)
No
Does a point-by-point radiofrequency ablation protocol targeting a proportional average impedance drop (%AID) ≥9% using the TactiFlex Sensor Enabled catheter achieve high first-pass isolation and durable pulmonary vein isolation at mandatory remapping in patients with atrial fibrillation?
This prospective study provides the first invasive remapping validation that guiding high-power short-duration RF ablation with a filtered proportional average impedance drop (%AID ≥9%) and strict inter-lesion distance yields a 93.4% durable pulmonary vein isolation rate without protocol-related complications.
Answers are grounded in this paper and cite their sources.
Supports %AID-guided ablation for durable PVI in AF; hypothesis-generating and leaves open randomized confirmation.
Background Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation (RF) catheter to guide pulmonary vein isolation (PVI) are limited. The use of impedance drop correlates with lesion formation but is hampered by cardiac and respiratory oscillation artifacts. A filtered impedance signal, average impedance drop (AID), is now available in the EnSite X EP system. The study aim was to evaluate the efficacy and safety of a novel ablation protocol based on proportional (%)AID using the TFSE catheter, including mandatory remapping to assess PVI durability. Methods Pulmonary vein isolation with the TFSE was performed in patients with atrial fibrillation (AF) undergoing first-time PVI. The ablation protocol consisted of reaching %AID ≥9 or reaching maximum allowed RF time of 20 s and 10 s anteriorly and posteriorly. Protocol-mandated remapping was performed ≥2 months following index ablation. Results The study included 26 patients (paroxysmal/persistent AF = 17/9). Protocol adherence was 93%, and first-pass isolation was achieved in 85% of left and 93% of right pulmonary veins (PVs). Mean %AID was 9.4 ± 0.98 and proportion of lesions with %AID ≥9 was 62.2 ± 15.5%. Durable PVI was observed in 93% of all PVs at remapping (88% of patients) with no ablation-related complications. Inter-lesion distance, but not %AID, was significantly different in reconnected PV segments compared to segments with durable isolation (Firth logit <0.001). Conclusion Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.
No takes yet. Share an insight, caveat, or question.
Akerström et al. (2026) conducted an observational in Paroxysmal or persistent atrial fibrillation undergoing first-time pulmonary vein isolation (n=26). Proportional average impedance drop (%AID ≥9)-guided high-power short-duration (50 W) radiofrequency ablation using the TactiFlex Sensor Enabled (TFSE) catheter was evaluated on Pulmonary vein isolation (PVI) durability at protocol-mandated remapping (≥2 months post-ablation) and rate of first-pass isolation. Point-by-point RF ablation guided by %AID ≥9 with the TFSE catheter achieved 84.6% patient first-pass isolation and 93.4% durable PVI at remapping without ablation-related complications.