Key result
Wavelet pre-processing with fast-ICA reconstructs noisy magnetocardiogram waveforms with a ~0.98 correlation to controls.
Why the study?
In independent component analysis of magnetocardiograms, low signal-to-noise ratios—such as those caused by surgically fixed sternal wires—reduce noise rejection accuracy and make determining the contraction dimension difficult.
Answers are grounded in this paper and cite their sources.
Demonstrates wire noise rejection in magnetocardiograms using wavelet transformation and fast ICA, indicating improved waveform recovery.
We present a magnetic noise rejection method using wavelet transformation (WT) and independent component analysis (ICA) for analyzing magnetocardiograms (MCGs). In the ICA, when the signal-to-noise (SN) ratio of the data is low, the accuracy of the noise rejection becomes low and the determination of the contraction dimension becomes difficult. For example, MCG has a very low SN ratio in patients who have a wire surgically fixed in their sternum. To solve this limitation, we have proposed a method to distinguish signal and noise components using Mahalanobis distance and WT as pre-processing for ICA. We carried out a simulation which attached a wire to a normal subject in order to show accuracy improvement of the noise rejection by the proposed method. The SN ratio of measured MCG was the -6dB. The correlation coefficient of waveform after noise rejection with control data (waveform without the wire noise) was calculated. As this result, it became 0.98 high value. The proposed method was effective for the wire noise rejection.
No takes yet. Share an insight, caveat, or question.
KOBAYASHI et al. (2013) studied Simulated sternal wire magnetic noise in magnetocardiography (n=5). Wavelet transformation (WT) pre-processing combined with fast independent component analysis (fast-ICA) and Mahalanobis distance vs. Fast-ICA alone and 0.5 Hz high-pass filter (HPF) + fast-ICA (evaluated against wire-free control data) was evaluated on Correlation coefficient of reconstructed magnetocardiogram waveform after noise rejection compared with wire-free control data. Wavelet transformation pre-processing combined with fast-ICA achieved a 0.98 correlation coefficient with wire-free control magnetocardiogram waveforms under a -6 dB signal-to-noise ratio.
Synapse has enriched 10 related papers. Consider them for comparative context: