Key result
Independent component analysis effectively reduces sternal wire magnetic noise during magnetocardiography.
Why the study?
In magnetocardiogram measurements, magnetic noise from sternal suture wire used after heart surgery interferes with the signal.
Answers are grounded in this paper and cite their sources.
ICA may reduce wire noise in MCG; leaves open diagnostic utility after sternotomy.
In magnetocardiogram (MCG) measurements, magnetic noise from the wire used to suture the sternum after heart surgery becomes a problem. Independent component analysis is an effective method of noise rejection. In this study, MCG measurements were made on a normal subject with the wire attached and without the wire. We performed signal processing by independent component analysis in order to reduce the effect of magnetic noise from the wire. Comparison of the waveforms after this signal processing with waveforms without the wire clearly showed that magnetic noise caused by the wire was reduced. This result clearly shows that independent component analysis is effective for the removal of magnetic noise from the wire.
No takes yet. Share an insight, caveat, or question.
KOBAYASHI et al. (2007) studied Magnetic noise from sternal suture wire in magnetocardiogram measurements (n=1). Independent component analysis (ICA) signal processing vs. Magnetocardiogram waveforms measured without the wire was evaluated on Reduction of magnetic noise from the wire in magnetocardiogram waveforms. Independent component analysis signal processing effectively reduced magnetic noise caused by sternal suture wire in magnetocardiogram measurements of 1 normal subject.