Direct-conversion cadmium zinc telluride (CZT) detectors have well-documented advantages over the sodium iodide (NaI(Tl))/photomultiplier cameras still standard in clinical gamma imaging, but low-cost, indigenously developed CZT systems are seldom characterised in detail. We characterise the detector physics of one such camera, a portable Small Area Imaging Gamma Camera (SAIGC; 44×44 pixellated CZT array, 2×2×5 mm elements, 88×88 mm field of view), and benchmark it against a commercial GE Infinia NaI(Tl) system for planar thyroid scintigraphy. Thirty patients referred for ⁹⁹ᵐTc-pertechnetate scintigraphy were imaged on both systems, with every SAIGC figure derived from raw list-mode data and compared with the published NEMA NU-1 specifications of GE Infinia. Energy resolution was 3.52% full width at half maximum at 140 keV, against 9.8% for GE Infinia. Separately, clinical pixel counts followed Poisson statistics on short timescales, with no detectable correlated counting from the electronics. Median acquisition time was 119.6 s per view. Bench source and phantom studies showed spatial discrimination, simultaneous dual-isotope imaging (¹⁵²Eu/⁹⁹ᵐTc, only 18 keV apart) and thyroid morphology preserved over a full 0–360° n, indicating single-photon emission computed tomography (SPECT) feasibility. The camera delivers these detector-level gains, alongside clinically interpretable images, at roughly one-tenth the cost.