Answers are grounded in this paper and cite their sources.
To compare the diagnostic performance and reproducibility of vertical root fracture (VRF) detection on cone-beam computed tomography (CBCT) between two complete viewing workflows, interactive picture archiving and communication system (PACS) workstation interpretation and smartphone review of transmitted screen-recorded videos, in vitro. Seventy-two unobturated single-rooted teeth (36 with mechanically induced VRFs, 36 intact) in a bovine rib model were rated on a five-category scale by three blinded observers (one endodontist and two oral and maxillofacial radiologists) using PACS and, after a 3-week washout, WhatsApp-transmitted videos on a smartphone. Unable-to-judge responses were infrequent (8/432 ratings). Smartphone-minus-PACS differences ranged from −8.3 to 0.0 percentage points for sensitivity, −11.1 to +11.1 for specificity, and −0.052 to +0.046 for the area under the receiver operating characteristic curve. Confidence intervals were up to 26 percentage points wide, and no difference was significant after false-discovery-rate adjustment. Cross-workflow weighted κ was 0.810–0.934, and inter-observer κ was 0.757–0.898 on PACS and 0.745–0.885 on the smartphone. An exploratory interaction analysis suggested a less favorable smartphone effect for the endodontist, specific to the included observers. Equivalence was not established. Smartphone video review may serve only as an adjunct for preliminary remote consultation.
No takes yet. Share an insight, caveat, or question.
Ghoncheh et al. (2026) studied this question.