Answers are grounded in this paper and cite their sources.
OBJECTIVE This study aimed to evaluate the accuracy of continuous glucose monitoring (CGM)–derived glucose values versus venous plasma glucose values during oral glucose tolerance testing (OGTT) in youths with obesity, their effect on glycemic classification, and whether deconvolution-based reconstruction improves agreement. RESEARCH DESIGN AND METHODS Twenty-nine youths with obesity underwent OGTT while wearing a FreeStyle Libre 2 sensor. Raw CGM values were temporally aligned with venous plasma glucose values and used to reconstruct plasma profiles via a first-order plasma–interstitial kinetic model and inverse filtering. Accuracy, agreement, consensus error grid, and glycemic classification were compared with collected venous plasma glucose values. RESULTS Temporally aligned CGM-derived values showed limited agreement with collected venous plasma glucose values (mean absolute relative difference [MARD] 12.96%; 51.72% of values within ±10 mg/dL; 20.06% of values differing by >25 mg/dL) and caused classification shifts. After reconstruction, MARD decreased to 5.55%, the proportion of values within ±10 mg/dL increased to 75.86%, almost all values fell within consensus error grid zone A, and classification shifts were reduced. CONCLUSIONS Deconvolution-based CGM reconstruction may support plasma-referenced, less invasive OGTT-based dysglycemic assessment.
No takes yet. Share an insight, caveat, or question.
Piona et al. (2026) studied this question.